Listing #513131 – For Sale GAMBRO CWP-100: WRO-112 Chem Water Purification
|
|
This Premium Listing has been seen 24 times and clicked on 237 times (as of 02:00 am edt)
 |
|
Updated: August 20, 2008
|
|
|
Condition: Good
|
Date of Manufacture: 1996
Serial number: 639
The unit has been used in a French hospital and has been under maintenance by Gambro during its entire lifetime.
The unit is located in France
Technical specs:
Minimum capacity in liters/minute at a pure water outlet pressure of
200 kPa:
+ 5°C: 7.0 l/min
+10°C: 8.0 l/min
+15°C: 10.0 l/min
+20°C: 12.0 l/min
Water Supply
Min. input: 20 l/min
Min. input pressure: 200 kPa
Max. input pressure: 500 kPa
Temperature: +5°C to +25°C (Maximum +20°C if peracetic acid is used for disinfection)
Quality of the purified water:
Depends on inlet water quality. If potable water is used and the
system is properly maintained, the following rejection rates will be
obtained:
Total dissolved salts: > 95%
Bacteria (CFU) & endotoxins: > 99%
Water conversion factor: max. 75%
Water quality of inlet water (demands):
Potable water should be used. Additional pretreatment is normally
necessary. Membrane and machine life expectancy depend on inlet
water quality. It is recommended not to operate the CWP outside the
following limits:
Hardness < 1 °dH
Iron < 0.1 mg/l
Manganese < 0.1 mg/l
Chloride < 100 mg/l
Silica < 25 mg/l
Total dissolved salts < 1500 mg/l
Chlorine < 0.1 mg/l
Fouling index ( S.D.I.) < 5
Drain Requirements
Required capacity: 20 l/min
Maximum flow occurs during flushing.
Membrane material: Modified polyamide, thin film composite
Membrane configuration: Spiral wound
pH-tolerance: 2 – 11
Power Supply
Mains voltage: 230/400 V, 50 Hz
Power rating: RO/RO-unit: 1.8 kW
Fuse: 10 AT
Dimensions
Weight: 225 kg.
Depth (mm): 670 (including handle, 40 mm)
Width (mm): 700
Height (mm): 2010
500 mm extra space is required on both sides of the unit.
|
Lister
| Anders Boersen Nielsen |
AGITO Medical |
| Phone: +33 686 685 967 | Country: France |
| |
|
DOTmed's COMPLIANCE NOTICE AND COMPLIANCE COMPLAINT PROCEDURE.
|