Listing #515582 For Sale GAMBRO CWP-100: WRO-101 H Water Purification
GAMBRO CWP-100: WRO-101 H Water Purification for sale   
Updated: August 20, 2008
Condition: Good
Date of Manufacture: 2002
Serial number: 1997
The unit has been used in a Danish hospital and has been under maintenance by Gambro during its entire lifetime.
The unit is located in Denmark

Technical specs:
Minimum capacity in liters/minute at a pure water outlet pressure of
200 kPa:
+ 5°C: 8.0 l/min
+10°C: 9.0 l/min
+15°C: 11.0 l/min
+20°C: 13.0 l/min

Water Supply
Min. input: 20 l/min
Min. input pressure: 100 kPa
Max. input pressure: 500 kPa
Temperature: +5°C to +25°C (Maximum +20°C if peracetic acid is used for disinfection)

Quality of the purified water:
Depends on inlet water quality. If potable water is used and the
system is properly maintained, the following rejection rates will be
obtained:
Total dissolved salts: > 95%
Bacteria (CFU) & endotoxins: > 99%
Water conversion factor: max. 75%

Water quality of inlet water (demands):
Potable water should be used. Additional pretreatment is normally
necessary. Membrane and machine life expectancy depend on inlet
water quality. It is recommended not to operate the CWP outside the
following limits:
Hardness < 1 °dH
Iron < 0.1 mg/l
Manganese < 0.1 mg/l
Chloride < 100 mg/l
Silica < 25 mg/l
Total dissolved salts < 1500 mg/l
Chlorine < 0.1 mg/l
Fouling index ( S.D.I.) < 5

Drain Requirements
Required capacity: 20 l/min
Maximum flow occurs during flushing.

Membrane material: Modified polyamide, thin film composite
Membrane configuration: Spiral wound
pH-tolerance: 2 – 11

Power Supply
Mains voltage: 380/400 V, 50 Hz
Power rating: RO/RO-unit: 1.8 kW
Heating unit: 7.0 kW
Fuse: 10 AT + 16 AT (slow blow)

Dimensions
Weight when shipped: 775 kg.
Depth (mm): 670 (including handle, 40 mm)
Width (mm): 1650
Height (mm): 2010
500 mm extra space is required on both sides of the unit.
Effective volume in heating tank: 260 litres



 Lister

Anders Boersen Nielsen AGITO Medical
Phone: +33 686 685 967Country: France
 

The lister has indicated that this item is no longer available.

DOTmed's COMPLIANCE NOTICE AND COMPLIANCE COMPLAINT PROCEDURE.

Access and use of this site is subject to the terms and conditions of our LEGAL NOTICE & PRIVACY NOTICE
Property of and Proprietary to DOTmed.com, Inc.
Copyright ©2001-2008 DOTmed.com, Inc.
ALL RIGHTS RESERVED