SN:P55132-K6
Leads: I, II, III
Display sensitivity: 0.25, 0.5, 1.0, 2.0, 3.0, 4.0 cm/mV ± 10%
Frequency response to screen: ESU filter out: 0.5 to 40 Hz, -3db max @ 1 mVpp input; ESU filter in: 0.5 to 20 Hz -3db max @ 1 mVpp input
Heart Rate Meter
Range: 30-300 bpm adult/child. 30-350 bpm neonatal
Accuracy: 30-250 bpm ± 3bpm or 3%, whichever is greater.
Pacer rejection: rejects all pulse of amplitude ± 2.0 mV to ± 700 mV and duration 0.1 ms to 2 ms with no tail. Rejects all pulses ± 2.0 mV to ± 700 mV and duration 0.1 ms to 2 ms with 100 ms T.C. tail <0.8 mV or 25 ms T.C. tail <0.4 mV.
Display update: 2 seconds ± 0.1 sec
NIBP
Technique: oscillometric
Systolic range: adult/ped mode: 40-250 mmHg, neonate mode: 15-150 mmHg.
Diastolic range: adult/ped mode: 30-200 mmHg, neonate mode: 10-150 mmHg.
Systolic accuracy: mean error less than 7 mmHg, standard deviation less than 8 mmHg (adult/pediatric). Mean error less than 3 mmHg, standard deviation less than 13 mmHg (neonatal).
Diastolic accuracy: mean error less than 7 mmHg, standard deviation less than 8 mmHg (adult/pediatric). Mean error less than 4 mmHg, standard deviation less than 7 mmHg (neonatal).
Heart rate meter range: 30-250 bpm.
Heart rate meter accuracy: ±3 bpm or ±3%, whichever is greater.
Pulse Oximetry
Accuracy (1standard deviation): 70-100%, ±2%; 60-69%, ±4%; 0-59%, unspecified.
Response time: 13 seconds to 95% of final step change % SpO2 value from 60-95% at 72 bpm
Recorder (optional)
Speed: 50, 25, 12.5, 6.25, 3.125 mm/sec ±5%
Note: 6.25 and 3.125 are for respiration only
Chart size: 50 mm ±2mm
Battery Requirements