FEATURES:
Intra-Aortic Balloon Pump
ECG Triggers
Pacer Triggers
Advanced pneumatic design
Permanently attached doppler
Detailed monitoring
Rotating monitor
SPECIFICATIONS:
Annotations:
ECG: Lead
Pressure: Scale
Parameters: heart rate, systolic, diastolic, mean, augmented pressure
Trend: Same as trend display
Recording Format:
Recorder: thermal array type recorder/2" thermal paper
Waveform Presentation: ECG, arterial pressure, optional balloon pressure waveform.
Graphic Presentation: heart rate trend, pressure trends.
Vent marker: trigger point, Inflate interval
Trend Display:
Memory: 8 hours total; 30 minutes of 5 minutes average; 7.5 hours of 15 minutes average.
Data Available: heart rate, peak systolic, pressure, diastolic pressure, mean and augmented pressure
Inputs/Outputs:
ECG Connectors: 6 pin 5-lead AAMI proposed standard; patient input 3-conductor phone jacks; monitor level.
Pressure Connectors: 6-pin male connector, Datascope-specified compatible transducer; 3-conductor phone jack; monitor level
Voltage Interface: ECG=1V/mV nominal arterial pressure=1V/100mmHG nominal
Pressure:
Pressure sensitivity to trigger: 15 mmHg min pulse amplitude
Transducer Requirements:
Sensitivity: 5.0microV/V/mmHG.
Excitation: +5V DC.
Zero Range: +/-120mmHg
Trigger Source:
ECG trigger requirement: 120microV +/- 20microV/40microV at max. gain.
Pacer Rejection: (no tails): +/- 2mV to +/- 700mV duration 0.1 ms to 2 ms.
Pressure: adjustable between 7 mmHg to 30 mmHg pulse amplitude.
Pacer A-V: A-V interval 80-225 ms rate < or equal to 125bpm.
Pacer A: trigger requirement (same as ECG).
Pacer V: <185bpm.
Internal: 40-120 in 5bpm increments; default rate 80 +/- 1 bpm
ESIS: automatic suppression with internal ECG amplifier
ECG:
Leads: I, II, III, avr, avl, avf, V
Frequency Response: 0.5 - 135 Hz output jack 0.5 - 12 Hz display
Leakage Current: <10microA source current, <20microA sink current.
Defibrillation Protection: discharge levels up to 500 joules.
Recovery Time: trace returns to screen, 2 sec. maximum
Display:
Type: Gas Plasma, 5.2" x 8.25", 132mm x 210mm.
Channels: ECG, Arterial Pressure, Optional balloon pressure waveform
Reference Line: 0-300 mmHg/max
Pneumatics:
System Compressor: Dual head diaphragm pump brushless DC motor.
Patient Balloon Pump Gas: Medical-Grade Helium
Blood Detection: Automatic protection of pneumatics by optical detection and isolation of blood in auto fill lint.
Electrical:
Battery Pack: 24VDC, 18AH
Type: Sealed Lead Acid, maintenance free
Battery Operating Time: 1.5 hours@150bpm.
Battery Recharge Time: 18 hours maximum from completely discharged to full.
AC Power Requirements: 500 VA normal, 90-130V, 50/60Hz; 200-260, 50-60Hz
Mechanical:
Dimensions: 48" H x 33" W x 19" D (104cm x 71cm x 61 cm)
Weight: 175lbs (80kg) without helium tank.
|
|