2003 GE Ovation Open .35T
CLINICAL APPLICATION Whole body
CONFIGURATION Open MRI
SURFACE COILS Integrated transmit and receive body coil;; head coil, CTL array, shoulder array, large extremity array, small extremity array
SYNCHRONIZATION Standard cardiac gating, ECG/peripheral, respiratory gating
PULSE SEQUENCES: SE, IR, 2D/3D GRE and SPGR, 2D/3D TOF, 2D/3D FSE, 2D/3D FGRE and FSPGR, SSFP, FLAIR, EPI, fat/water separation, single shot diffusion EPI
IMAGING MODES Localizer, single slice, multislice, volume, fast, POMP, multi slab, cine, slice and frequency zip, extended dynamic range, tailored RF
TR 1.3 to 12000 msec in increments of 1 msec
TE 0.4 to 2000 msec in increments of 1 msec
SINGLE/MULTI SLICE Simultaneous scan and reconstruction;; 80 images/second reconstruction
FOV 3cm to 40 cm continuous
SLICE THICKNESS 2D: 1.4mm - 20mm 3D: 0.2mm - 20mm
SPATIAL RESOLUTION 0.08 mm
MAGNET TYPE Permanent
BORE DIAMETER
or W x H 175 x 85 x 447 cm
MAGNET WEIGHT 8256 kg
H*W*D 530 x 175 x 250 cm
POWER REQUIREMENTS 200 - 480, 3-phase
COOLING SYSTEM TYPE Liquid helium
CRYOGEN USE 0.03 L/hr, holds 300 L
FIELD STRENGTH 0.35 T
STRENGTH 15 mT/m
5-GAUSS FRINGE FIELD 2.5 m/2.5 m
Return Policy