Service For GE HEALTHCARE Repairing Logiq e. Repairing Logiq e.

Faulty Phenomenon Malfunction: The cursor on the screen system did not respond to movements of the trackball horizontally (left and right), which complicated the control of the device. ... view more

June 03  

Service For GE HEALTHCARE Repair Voluson E10 BT16. Repair Voluson E10 BT16.

Faulty Phenomenon Malfunction: the issue has a progressive nature, that is, earlier during the scanning process, an error occurred sporadically with the frequency of "Error: Aurora communication... view more

June 03  

Service For GE HEALTHCARE Updating the software version and iron on the Voluson E8 device. the Voluson E8 device

Faulty Phenomenon The procedure for updating the software version and hardware on Voluson E8 (versions BT10 to BT13) has been carried out. To install the software version BT13, a replacement BackEnd... view more

June 03  

Service For GE HEALTHCARE Updating the software version and iron on the Voluson E8 device. the Voluson E8 device

Faulty Phenomenon The procedure for updating the software version and hardware on Voluson E8 (versions BT10 to BT13) has been carried out. To install the software version BT13, a replacement BackEnd... view more

June 03  

Service For GE HEALTHCARE Repair device Voluson E8 BT13.5. Repair device Voluson E8 BT13.5.

Faulty Phenomenon Malfunction: when connecting any sensor to any connector of the sensor (error PSS) solution: repair of the sensor connection plate RTF was carried out. Troubleshooting and... view more

June 03  

Service For GE HEALTHCARE Repair Beamformer of Voluson E10 BT16. Repair Beamformer of Voluson E10 BT16.

Faulty Phenomenon Beamformer consists of RFM201, RFM221 (PN: KTZ303915, KTZ303916); RSX10, RSX20 (KTZ303250, KTZ303054). Error: Aurora communication failed; when turning on the device, the image... view more

June 03  

Service For GE HEALTHCARE Repair Voluson E10 BT21. Repair Voluson E10 BT21.

Faulty Phenomenon Malfunction: Memory access violation, device requires restart. Repair board RFM423G. P19 (PN: KTZ304074, KTZ304142): restoration of factory firmware in EEPROM and Flash Memory for... view more

June 03  

Service For GE HEALTHCARE Repair Logiq S8 R1. 1.1. Error Code: 0100 1.2. Error Description: Power Repair Logiq S8 R1. 1.1. Error Code: 0100 1.2.

Faulty Phenomenon Malfunction: artifacts on the display, improper operation of ATsP and TX Pulser (hdl6v5541) in the Beamformer BF192 Assy (PN: 5357234-2). Corresponding microchips have been... view more

June 03  

Service For GE HEALTHCARE Repairing Voluson E8 BT13. Repairing Voluson E8 BT13.

Faulty Phenomenon Malfunction: An artifact appeared in the form of a stripe on all sensors. After performing the service test BF, we detected a malfunction in the RSR board (KTZ196310, Receiver RX... view more

June 03  

Service For GE HEALTHCARE Repair Logiq S8 (R2). Repair Logiq S8 (R2).

Faulty Phenomenon Malfunction: Artifacts appeared in the form of white stripes during operation with any sensor. During the diagnostic process, we detected a malfunction in the BF192 board (PN:... view more

June 03  

Service For GE HEALTHCARE Repairing Voluson S6. Repairing Voluson S6.

Faulty Phenomenon Repair flat panel DRFG ASSY, DRFM-2 FRU SET, DRFM-3 ASSY (PN: S5814538, S5497712-2, S5497712-3) (also compatible with Voluson S8 and Voluson S10). Repair CPU (PN: 6324556)... view more

June 03  

Service For GE Logiq e Nextgen R8 Shared Service

Exchange - Send in your broken item and we will send you a working one of the same model while we... view more

June 03  
Bryan Carlin / CC Medical LLC
phone: +1 (833) 309-4229

Service For GE 1.5T 30ch AIR Coil REF: 7771501 FRU: 7771707 MRI Coil Repair Service MRI Coil

⚠️ REPAIR SERVICE LISTING – THIS IS NOT A COIL SALE MRI Coil Repair Service – Independent UK Specialist- ISO 13485 & ISO 9001 Certified Skylake Medical is a UK-based specialist provider of MRI... view more

May 11  

Service For GE C1-5-RS Extensive repair capability - C1-5-RS Ultrasound Transducer

Our company has 25 years' expertise in the ultrasound sector, 600 models of probes sold and/or repaired and 96% of satisfied customers. Certified "ISO 13 485", our company ensures not only a know-how... view more

June 03  
Laurent Souef / PRS HEALTHCARE
phone: +33 0472322626

Service For GE 12L-RS Extensive repair capability - 12L-RS Ultrasound Transducer

Our company has 25 years' expertise in the ultrasound sector, 600 models of probes sold and/or repaired and 96% of satisfied customers. Certified "ISO 13 485", our company ensures not only a know-how... view more

June 03  
Laurent Souef / PRS HEALTHCARE
phone: +33 0472322626