PHILIPS Panorama 0.23T defect
April 21, 2013 12:53

you can check receive cable grounding, if ok, try to replace mixer....Click here to read more

PHILIPS Panorama 0.23T defect
April 16, 2013 10:38

you'd better disconnect all the cables of not related scanning, and perhaps you should check gradient cable filters at RF panetration panel....Click here to read more

PHILIPS Panorama 0.23T defect
April 15, 2013 10:11

is there some equipment panetrate RF cage,like injector? try to disconnect light and service receptacle wires. check airconditioner duct....Click here to read more