Ray Brown - Freeman Health System - DOTmed Profile
advertisement
Current Location:
>
> Ray Brown


Ray Brown - Profile
BMET II
Freeman Health System
MO, USA
DOTmed User since May 2014
Email Ray

Rate This User
(30) (4)
Ray Brown MO USA BMET II

Ray Brown - Freeman Health System

Freeman Health System Joplin, MO

Overall Feedback Received

No ratings.

Recent Forum Posts

Phillip Intellvue MP 50
July 21, 2019 12:57

That would be if it said "NBP Equip Malf".....Click here to read more

Phillip Intellvue MP 50
July 19, 2019 10:45

There is no specific error that says exactly what you said. The word Occlusion does not appear as an Alarm or INOP. The only thing that comes close is if you go into the Setup of the NBP and select the VeniPuncture tab, which wil...Click here to read more

Zoll CCT Defibrillator Battery
April 06, 2019 12:31

No, there's no difference. The CCT option adds 2 invasive blood pressures, NIBP, temperature, and a remote VGA connector to the panel. Otherwise it uses the same batteries, either the SLA or the NiMH. Have you run the batteries th...Click here to read more