ge ct/i hispeed advantage
September 21, 2012 07:25

I hope you have already checked all your phases and rectified HVDC...However, When you did the man_hvdc test, you had LO SIDE (DS2 )not lit. That indicates Cathode. Now, I would guess Cath Inverter or tank as fault....Click here to read more

ge ct/i hispeed advantage
September 18, 2012 11:25

Then check brushes, slip ring. I believe the Anode and Cathode inverters and the rotor ctrl are using the 550v rail voltage. ...Click here to read more

ge ct/i hispeed advantage
September 16, 2012 10:43

with power off, I would disconnect the large capacitors(usually blue) and check their capacitance, compare. Should be 4600uf. You could also swap them & watch problem change. These caps balance the HV for anode/cathode. If this is...Click here to read more