Philips Respironics BiPAP Focus Features
The Philips Respironics BiPAP Focus is a non-invasive ventilator (NIV) that offers bi-level assistance ventilation to patients. This type of ventilation is specifically designed for facilities like hospitals. The assisted ventilation is suitable for patients with Obstructive sleep apnea, low acuity, and patients with respiratory insufficiency and distress. The BiPAP Focus is not a Critical care Ventilator. It is approved for acute, Sub-acute, and intra-hospital transport setting only. The BiPAP Focus comes with a standard alarm system and Auto-track Sensitivity that warn of triggers and different sensitivities that can change the breathing pattern of the patient and or cause ventilation leakage.
Support for patients weighing over 66 lbs (30kg)
Backup battery for continuous ventilation.
Portable design for intra-hospital transport.
Digital Auto-Trak adapts to patient changes.
Large user-interface.
Ideal for low-acuity environments.
Auto-adaptive triggering and cycling thresholds.
Philips Respironics BiPAP Focus Specifications
Dimensions
Height: 5.5” (14 cm)
Width: 11.4” (29 cm)
Depth: 14” (35.5 cm)
Max Weight: 10 lbs (4.5 kg)
Controls
CPAP: 4-20 cmH2O
EPAP: 4-25 cmH2O
IPAP: 4-30 cmH2O
Rise Time: 1-6 (1 = 0.1 s and 6 = 0.6 s)
Rate: 1-30 BPM (accuracy: ±BPM or ±10% of setting)
I-time: 0.5-0.3 s
Ramp Time: 0-45 min
Ramp Start: 4 cmH20 to EPAP
Maximum Flow Capacity: 160 l/min
Leak Compensation: Up to 60 l/min
Patient Parameters
Patient Circuit Pressure: 0-35 cmH2O
Rate Range: 0-99 BPM
Estimated VT Range: 0-4000 ml
Estimated MV Range: 0-99 l/min
Patient Leak Range: 0-200 l/min
Apnea Rate Range: 0-99 per hour
Alarms
High System Pressure: 5 cmH2O over target IPAP pressure
Low System Pressure: 5 cmH2O under target EPAP pressure
Apnea Range: 20, 40, 60 s or off
Apnea Rate Range: 5, 10, 20 apneas per hour or off
Nurse Call: Normally open, normally closed
Electrical
Input Rating: 100 VAC / 230VAC
Battery Backup: 3.8 amp-hour (NiMH)
Fuse: No replaceable fuse
Nurse Call: 0.280 A (rated current); 250 V (rated voltage)
"The sale of this item may be subject to regulation by the U.S. Food and Drug Administration and state and local regulatory agencies. If so, do not bid on this item unless you are an authorized purchaser. If the item is subject to FDA regulation, I will verify your status as an authorized purchaser of this item before shipping of the item."
All units have been cleaned and handled in accordance with the manufacturer's instructions.
DISCLAIMER: Regardless of the origin of the equipment, documentation provided or identification appearing upon the equipment, the equipment described and offered here is in no way certified for, recommended for, or offered for any specific use. The purchaser agrees that the seller shall not be held responsible or liable for any injuries or damages, whether incidental or consequential, associated in any way with the equipment. The purchaser, by bidding on this equipment, indicates their acknowledgment of, and agreement to the terms of this disclaimer.
SHIPPING: Customer will be responsible for all shipping fees.
Return Policy
Items are sold as-is with no returns or refunds available unless explicitly stated.
Return Policy