From routine to specific tests, perform them all on a single system
*55 on-board assays
*Innovative cassette format
*Integrated work station and validation station
*Touch screen
*Real-time viewing of your workflow
General Description:
Analysis Method Types:
Spectrophotometry: Colorimetry and Turbidimetry
Potentiometry: Direct (Serum or Plasma) and Indirect (Urine)
Up to 55 assays on board
Optional ISE Module (Na+, K+, Cl-)
Maximum Throughput: 420 test/h
CE Marked
Sample Handling:
Sample Capacity: 6 racks with 10 sample positions for primary tubes or cups
Continuous loading
Positive Sample Identification: Integrated Barcode Reader and Tube Presence Detector
Sample Integrity: Level, Shock and Clot Detection
Sample Collection Format: 0.7, 4, 5, 7 and 10 mL (diameter: 16mm)
Sample Specimen: Serum, Plasma, Urine, CSF, Whole Blood, Homogeneous Fluids
Sample Volume: 2 to 380 µL
Automatic Sample Dilution: from 1/2 to 1/22,500
ISE: 20 µL (Indirect) or 60 µL (Direct)
Automatic Post Concentration: from x2.0 to x 10.0
Automatic Re-run
Calibrator/Control Handling:
Calibrator/Controls Positioning on Sample or Reagent Tray
Automatic Standard Dilution Series:
Dilution Factors: Main Direct 1/2 - 1/150
Dilution Factors: Main Indirect 1/2 - 1/1508
Reagent Handling:
52 Reagent positions on board:
44 in a closed refrigerated area (2-10°C)
8 at room temperature
Positive Reagent Identification:Integrated Barcode Reader
Reagent Probe : Level Detection and Preheating
Reagent Management with Cassettes: Automated Reagent On-board Stability Check plus Test Countdown with automatic back-up
Reagent Volume: from 2 to 600 µL
Reaction System:
Reaction Volume: 150 to 600 µL
Reaction Temperature: Air bath at 37°C
Photometer Reaction Cycle:12 seconds
Mixing with Teflon-coated Stirrer
Reaction Time: 1 to 100 cycles
Disposable Acrylic Cuvettes
Automatic Cuvette Changer
Cuvette Capacity: 36 segments (432 cuvettes)
7 Integrated Calculation Algorithms