Mammo GE Senocare Essential
Inspection done by & Dare: Laurent MICAN, ALTHEA 23 February 2023
Manufacturer and model: GE ESSENTIAL
System Yom: 07-2012
Tube Type: APOLLON
Tube Yom: 02-2020
Tube installation date: 08-2020
Workstation model no.: CONTROL STATION V4
Software
Manuals
Functional condition: CALIBRATION LE 13/02/2023, PASS
.
TESTS NAMES PERFORMED:
1. Gain Detector Calibration MO/MO-GRAID-LF (PASS)
2. Gain Detector Calibration MO/Rh-GRAID-LF (PASS)
3. Gain Detector Calibration RH/Rh-GRAID-LF (PASS)
4. Gain Detector Calibration MO/MO-NOGRID-LF (PASS)
5. Gain Detector Calibration MO/Rh-NOGRID-LF (PASS)
6. Gain Detector Calibration RH/Rh-NOGRID-LF (PASS)
7. Gain Detector Calibration MO/MO-NOGRID-SF (PASS)
8. Gain Detector Calibration MO/Rh-NOGRID-SF (PASS)
9. Gain Detector Calibration RH/Rh-NOGRID-SF (PASS)
10. Tube tilt Calibration (PASS)
11. Collimator (PASS)
12. Brightness Non Uniformity (PASS)
13. High Frequency Modulation (PASS)
14. Bad Pixel (PASS)
15. Bad Roi (PASS)
16. SNR Non Uniformity (PASS)
Return Policy
Items are sold as-is with no returns or refunds available unless explicitly stated.
Tags:
Mammography