For Sale VALLEYLAB Force FXc Electrosurgical Unit

  • Asking Price :$3,750.00 USD [convert]
  • Condition : Refurbished
  • Weight : 18.00 Lbs
  • Qty. Available : 1
  • In Stock : Yes
  • Date : May 10, 2025
  • Listing # : 1174052
cps-logo Get a CPS 2-Year Protection Plan starting at $5.99. Sold separately. Learn more
Advertisement

Item Description

Seller assumes all responsibility for this listing..

Listing: #1174052

  • Condition: Refurbished
  • Brand: VALLEYLAB
  • Type: Electrosurgical Unit
  • Model: Force FXc
  • Warranty: 6-MONTHS

REFURBISHED / COVIDIEN Valley Lab Force FXc
Unit has been bio-medically inspected and tested. Performs to OEM specifications.

FEATURES

•Instant Response™ technology ensures that the power delivered remains virtually constant, regardless of the tissue type
•Improved performance at lower power settings minimizes the risk of tissue damage and neuromuscular stimulation
•Three internal microcontrollers reduce system reaction time and increase the system’s processing speed
•Spray coagulation voltage of no more than 9000 volts peak-to-peak output for broad, but superficial coagulation with limited capacitive coupling
•A Power Efficiency Rating (PER) of approximately 98 for accurate and consistent cut performance
•Three cut modes, all controlled by Instant Response™ technology, offer surgeons a variety of choices
- Low Cut for delicate tissue or laparoscopic cases
- Pure Cut for a clean, precise cut
- Blend for cutting with hemostasis
•Four coag modes
- Desiccate for low voltage contact coagulation suitable in laparoscopic and delicate tissue work
- Fulgurate (high crest factor) for efficient noncontact coagulation in most applications
- Fulgurate (low crest factor) for lower voltage coagulation requirements
- Spray for coagulating large tissue areas with superficial depth of necrosis
•Three bipolar modes
- Precise, Standard, and Macrobipolar are controlled by the Instant Response™ system
- Precise and Standard setting utilize low voltage to prevent sparking
•Versatile system that is uniquely compatible with other devices, including:
- Force Argon™ II and Force GSU™ argon coagulation system
- CUSA EXcel™ and CUSA™ 200 ultrasonic surgical aspirators
- OptiMumm™ smoke evacuator, through a direct cable link
- Valleylab VLCM bipolar current monitor
•Compatible with and used as the electrosurgical energy source for:
- Dyonics* Control RF arthroscopic ablation system
- Dyonics* Electroblade™ rotary resection system
- Cook Vascular Perfecta™ EDS pacemaker lead extraction system
•Compatible with, and the exclusive electrosurgical generator for, the Computer Motion Hermes™ Voice Command System
* Trademark of Smith & Nephew


TECHNICAL SPECIFICATIONS (110-120V)

Output Waveforms

Bipolar
Precise: 470 kHz sinusoid
Standard: 470 kHz sinusoid
Macro: 470 kHz sinusoid

Monopolar Cut
Low: 390 kHz sinusoid. Similar to the Pure Cut mode except the maximum voltage is limited to a lower value.
Pure: 390 kHz sinusoid
Blend: 390 kHz bursts of sinusoid, recurring at 27 kHz intervals. 50% duty cycle envelope.

Monopolar Coag
Desiccate: 240 kHz sinusoid repeated at 39 kHz. 8% duty cycle.
Fulgurate: 390 kHz damped sinusoid bursts with a repetition frequency of 30 or 57 kHz into 500 ohms
Spray: 390 kHz damped sinusoidal bursts with a randomized repetition centered at 28 kHz. Frequencies include 21 kHz
Output power changes by less than 15% or 5 watts, whichever is greater, as the line voltage varies from 104-132 volts and 208-264 volts (at rated load).

Low Frequency Leakage (50-60 Hz)
Source current, patient leads, all outputs tied together

•Normal polarity, intact chassis ground: <10 µA
•Normal polarity, ground open: < 50 µA
•Reverse polarity, ground open: < 50 µA
•Sink current, 140V applied, all inputs: < 50 µA
PER = 98

PER (Power Efficiency Rating) is the measure of an electrosurgical generator’s ability to accurately deliver the selected power into a wide range of tissue types.

Weight and Dimensions

•Height: 11.1 cm (4-3/8 in.)
•Width: 35.6 cm (14 in.)
•Length: 43.9 cm (17 in.)
•Weight: < 8.1 kg (<18 lbs)
Input Power Requirements

Operating range is 85 to 132 AC volts. Maximum current is 7 amperes in Cut and 4 amperes in Coag.

High Frequency Leakage

•Bipolar: Less than 60 mA (rms)
•Monopolar: Less than 150 mA (rms)
REM™ Contact Quality Monitoring System

Measurement Frequency: 80 kHz ± 10 kHz

Measurement Current: Less than 10 µA

Acceptable Resistance Ranges:

•REM™ pad – 5-135 ohms
•Non-REM™ pad – less than 20 ohms
Acceptance range is 5-135 ohms after REM PolyHesive II return electrode is applied.

Adaptive REM™

REM™ trip is baseline impedance plus 40%. For example, if the baseline impedance is 30 ohms, the upper level trip approximately 42 ohms. If the pad-patient impedance falls below the baseline impedance, a new baseline is established.


OUTPUT CHARACTERISTICS

Bipolar

Precise Mode:
Maximum P-P Voltage: 450
Rated Load (OHMS): 100
Maximum Power (Watts): 70
Crest Factor* (Typical): 1.5

Standard Mode:
Maximum P-P Voltage: 320
Rated Load (OHMS): 100
Maximum Power (Watts): 70
Crest Factor* (Typical): 1.5

Macro Mode:
Maximum P-P Voltage: 750
Rated Load (OHMS): 100
Maximum Power (Watts): 70
Crest Factor* (Typical): 1.5

Monopolar Cut

Low Mode:
Maximum P-P Voltage: 1350
Rated Load (OHMS): 300
Maximum Power (Watts): 300
Crest Factor* (Typical): 1.5

Pure Mode:
Maximum P-P Voltage: 2300
Rated Load (OHMS): 300
Maximum Power (Watts): 300
Crest Factor* (Typical): 1.5

Blend Mode:
Maximum P-P Voltage: 3300
Rated Load (OHMS): 300
Maximum Power (Watts): 200
Crest Factor* (Typical): 2.5

Monopolar Coag

Desiccate Mode:
Maximum P-P Voltage: 3500
Rated Load (OHMS): 500
Maximum Power (Watts): 120
Crest Factor* (Typical): 5

Fulg. High Crest Factor* Mode:
Maximum P-P Voltage: 8500
Rated Load (OHMS): 500
Maximum Power (Watts): 120
Crest Factor* (Typical): 7.0

Fulg. Low Crest Factor *Mode:
Maximum P-P Voltage: 6900
Rated Load (OHMS): 500
Maximum Power (Watts): 120
Crest Factor* (Typical): 5.5

Spray Mode:
Maximum P-P Voltage: 9000
Rated Load (OHMS): 500
Maximum Power (Watts): 120
Crest Factor* (Typical): 8

CEM™ Mode


Monopolar Cut (Low):
Maximum P-P Voltage: 1000
Rated Load (OHMS): 300
Maximum Power (Watts): 100
Crest Factor* (Typical): 1.5

Monopolar Coag (Dessicate):
Maximum P-P Voltage: 3500
Rated Load (OHMS): 500
Maximum Power (Watts): 70
Crest Factor* (Typical): 5

* Crest Factor is an indicator of a waveform’s ability to coagulate without cutting.