Get a CPS 2-Year Protection Plan starting at $5.99. Sold separately. Learn more
Seller assumes all responsibility for this listing..
Listing: #5624253
User-configurable visual and audible display and alarm.
Controllable from inside and outside the scanner room with a remote keypad.
Portable, stand-mounted monitor.
Modular platform that can be customized and updated to meet your needs.
Monitors agent gases, pulse oximetry, ECG, NIBP, Invasive Pressure, CO2.
Dimensions
Height: 8.5” (21.6 cm)
Width: 18.6” (47.7 cm)
Depth: 13.4” (34.2 cm)
Weight: 38 lb (17.2 kg)
Display
Type: Backlit electroluminescent
Size: 7.7 x 4.8” (19.5 x 12.2 cm)
Resolution: 640 x 200
Strip Printer
Print Speed: 12.5, 25, or 50 mm/sec for cardiac waveforms 3, 6, 12, or 25
Waveforms: Prints up to four groups of waveforms with up to 2 waveforms per group
Continuous: Prints a continuous strip chart of group 1 waveforms, with annotation lines printed every 200 mm (8”)
Anesthetic Agents
Technique: Piezoelectric Crystal Resonance
Gases: Isoflurane, Enflurane, Halothane, Desflurane, Sevoflurane
Display Range: 0 to 20%
Accuracy: 10% or 0.2%, whichever is greater
Oximetry
SpO2: Range: 0 – 100%; Accuracy: ±2% at 70%; ±3% at 50% – 69%; Calibration: Fractional
Pulse Rate: Range: 30 to 254 BPM; Accuracy: ±2% at 30 – 100 BPM
ECG
Heart Rate: Range: 30 – 250 BPM; Accuracy: ±2% at 30 – 100 BPM
Lead Selection: I, II, III
Bandwidth: 0.5 Hz to 40 Hz
%O2
Range: 0% to 100% O2
Accuracy: ±2% O2 (When % O2 cell is calibrated with 100% O2)
Resolution: 1% O2
Probe: Galvanic Fuel Cell
NIBP
Range: 10 – 300 mmHg
Accuracy: ±4 mmHg
Inflate Time: 10 seconds
Invasive Blood Pressure