Technical parameters
Imaging system
Anatomical regions: regions consistent with studies
using this system are: calf, knee, ankle, foot,
thigh, hand, wrist, forearm, elbow, arm,
shoulder, hip, lumbar-sacral and cervical
column, temporomandibular joint (sublicense)
Excited nuclei: lH
Acquisition and reconstruction methods:
- Multidimensional 2D acquisition, with 2D Fourier transform reconstruction,
BMTEC/23/12/2022
with zero fill for partial acquisitions and
for rectangular FOV
- Multilayer 2D, half-echo
- Multilayer 2D, half-scan
- Multilayer 2D, turbo
Acquisition 3D, with reconstruction according to
Fourier transformed 3D (in isotropic 3D
the reconstructed volume can be stored both
on hard disk and optical disk for subsequent
multiplanar relicing, in anisotropic 3D a
multilayer is produced as in 2D);
- 2D in real time as a reference image for
positioning
- dynamic acquisition in given time
intervals
Acquisition matrix: 2-D: from 128x128 to 512x512; phase
encoding step is 8, and frequency
encoding is 32
3D: from 192x128 to 256x256; slice
encoding from 24 to 128, step 8; phase
encoding step is 8
Spatial resolution: up to 0.2 mm nominal
Field of view (FOV): 100 mm to 400 mm in 10 mm steps, to
manage signal-to-noise ratio and image resolution
Usable and display area (maximum
displayed image) corresponds to maximum
FOV 270 mm
Number of slices x echoes: 1 to 128 (max. 2 echoes), in case of
Turbo acquisitions (with several echoes)
limit is adjusted accordingly (min. 16) •
Slice selection method: selective gradient and RF pulse
Slice thickness: 2-D: 2 mm to 10 mm in 0.5 mm steps
3-D: 0.6 mm to 10 mm in 0.1 mm steps
Inter-slice distance: adjacent slices can be obtained
BMTEC/23/12/2022
Slice orientation: transverse, sagittal, coronal, oblique
Packages: multiple parallel and non-parallel packages
TR regardless of the number of slices
Magnetic system
Magnet type: permanent, in NeFeB, proprietary type 'C'
Magnetic circuit
Field direction: vertical, between two C walls
Static field strength: 0.24 T ± 5.0 mT (f0: 10.22 MHz ± 210 KHz)
Thermal control: 5 independent channels for controlling
magnet temperature at 35 °C
Field stability: <30 Hz/10 minutes
Uniformity: ± 3 ppm FWHM over 270 mm DSV
Field stability: 30 Hz/10 minutes
Pad system: passive
Field edge (0.5 mT line): contained at a distance of 1.8 m from the center of the magnet
Magnetic shielding: not necessary
Gate opening: Net height of the shield recess: 37.5 cm
Height of the recess from the patient table surface,
without cushions: 35.1 cm
Height of the recess from the patient table surface,
with cushions: 32.3 cm
Cushion thickness: 2.8 cm
Dedicated gradient system
Coils: 3 gradient coils (X, Y, Z), one for
uniform field (Bo)
Maximum intensity: ± 20 mT/m
Rise rate: 56 mT/m/ms
Technical description
Linearity: ± 5% per 200 mm DSV; ±12% at 230 mm
DSV
Cooling system: both coils and amplifier are air cooled
Control system: digital based on SHARC 21161 DSP
(600 MIPS, 600 MFLOPS, 128 KB)
BMTEC/23/12/2022
4 independent channels (X - Y - Z - Bo)
18 bit DAC - update every 7.2 µs
Ramp generation - induced current
compensation pre-emphasis - adjustable
delay
Calibration procedure: all calibrations of the gradient system are
managed by the console at different levels
automation
Magnetic compensation system (optional):
'open loop' control of magnetic changes
field via external measurement (AC and DC
sensors), digital signal processing in
separate channels (DC - 50/60 Hz - 16.6 Hz)
and corrective control by means of a Bo coil.
Digital electronics based on SHARC 21161 DSP
Radio frequency, transmission and acquisition system
Receive coils: Shoulder coil 1 (linear): 22.4 x 21.6 x 12.6
(ext.) 14.5 x 17.5 x 12.6 (in.) cm
Knee coil 2 (DPA): 22.5 x 21.0 x 18.3 (ext.)
14.3 x 16.0 x 18.3 (in.) cm
Hand/wrist coil 3 (DPA): 17.8 x 17.5 x 20
(ext.) 11.9 x 7.2 x 20 (in.) cm
Ankle coil 4 (DPA): 22.0 x 19.2 x 28.5 (ext.)
14.6 x 10.0 x 28.5 (in.) cm
Flexible coil 6 (linear): 4.0 x 33.0x x 28.5 cm
(maximum inner)
Bracelet coil 7 (DPA): 21.5 x 30.0 x 21.8
(maximum inner)
Neck coil 9: 25.5 x 33.0x x
33.0 cm (maximum outer)
Lumbar coil 17 (multi-channel): 27.0 x 28.0
x 40.0 (inner) cm 30.5 x 28.0 x 48.8 (outer) cm
Lumbar coil 18 (multi-channel): 28.8 x 28.0
x 47.0 (inner) cm 32.3 x 28.0 x 55.4 (ext.) cm
BMTEC/23/12/2022
Optional receiver coils: Coil 14 lumbar-sacral (DPA): 29.4 x
32.2 x 42.3 (max. ext.), surface per limb: max.
height 22.1 cm, min. height 20.6 cm, width 19 cm, length 13.1 cm
TMJ coil 15: max. height 23.7 cm, max.
width 14.4 cm, thickness 2.8 cm Gantry: includes gradient and RF coils
transmit coil; allows quick positioning and connection
with quick connector on both sides
opening and automatic recognition
receiving coil
Receiving chain: 4 independent channels
noise figure <1 dB
bandwidth at 0.5 dB: 600 KHz (10.2 MHz ±300 kHz)
maximum power 1500 W
Power supply
Mains parameters: 200÷210/220÷240 VAC, 50/60 Hz, 24 hours
Consumption: 2100 VA during fast heating
2400 VA in acquisition
1500 VA in rotation
1200 VA in standby mode
400 VA with unit OFF (thermal control)
Current protection: thermal switch 16 A primary, 16 A secondary
Computer and monitor consumption: 400 VA
Grounding: via 6 mm2 power cable Overall dimensions and weight
Magnet: H (Y direction in vertical field) approx.
145 mm; W (X direction horizontal) approx.
149 mm; D (Z direction horizontal) approx.
105 mm
Weight approx. 6050 kg
Magnet mounted on rotating structure:
H (Y direction vertical in field) approx.
199 mm; W (X direction horizontal) approx.
211 mm; D (Z direction horizontal) approx.
BMTEC/23/12/2022
180 mm
Weight approx. 7455 kg
Bed: approx. 98 kg
Electronic box: 60x72x90 cm; 201 kg empty, 341 kg full
PC: 18x45x48 cm; 13 kg
PC packaging: 30x70x60 cm; 18 kg
24' TFT monitor: 55.68x37.78 ÷ 52.8x22.76 cm; 10.6 kg
24' TFT monitor in packaging: 65x50x32.5 cm; 14 kg
Return Policy