Hitachi Aperto Lucent – 0.4T MRI System (2016)
General Information
• Model: Aperto Lucent
• Year of Manufacture (YOM): 2016
• Software Version: V3.5D
• Condition: Excellent working & cosmetic condition
• Included: Software CDs, Phantom, Standard Accessories
Coils Included
1. Head Coil
2. Body Coil (Large)
3. Body Coil (Medium)
4. Knee Coil
5. Joint Coil (Medium Large)
6. Joint Coil (Small)
7. Shoulder Coil
________________________________________
System Highlights
• Permanent Magnet, Open Design: 0.4 Tesla mid field MRI with single pillar gantry and 320° accessibility.
• Patient Comfort: Side loading table, quiet operation, and open bore design reduce anxiety and improve workflow.
• Advanced Motion Correction: RADAR technology enhances imaging quality for musculoskeletal and general applications.
• Cost Effective: Lower installation and operating costs compared to superconducting systems.
________________________________________
Key Technical Specifications
Feature Specification
Magnetic Field Strength 0.4 Tesla (T)
Magnet Type Permanent
Configuration Open Bore (Single Pillar, 320° opening)
Max. Gradient Amplitude 22–25 mT/m
Gradient Slew Rate 55 T/m/s
Maximum Patient Weight 225 kg (≈ 496 lbs)
Generator Power 2.5 kW
Power Supply Capacity 9.5 kVA
Minimum Room Size 30 m²
Table Features Motorized, height adjustment, lateral movement ±150 mm
________________________________________
Imaging Features
• Modes: T1, T2, DWI, FLAIR, MRCP, MRA (2D/3D TOF)
• Fat Suppression: CHESS RF saturation sequences
• Shimming: 8 channels of active patient shimming for uniformity
• Coil Technology: Multi channel RF coils with solenoid receiver design for high SNR and resolution
• Software & Workflow:
o IP Recon iterative reconstruction algorithm
o SynergyDrive technology for enhanced efficiency
Return Policy