Over 450 Total Lots Up For Auction at Three Locations - NY 05/20, TX 05/22, WI 05/27

spacer
ENVITEC Compatible O2 Cell for Covidien Puritan Bennett - 4-072214-00 Oxygen Sensor ENVITEC Compatible O2 Cell for Covidien Puritan Bennett - 4-072214-00 Oxygen Sensor ENVITEC Compatible O2 Cell for Covidien Puritan Bennett - 4-072214-00 Oxygen Sensor
ENVITEC Compatible O2 Cell for Covidien Puritan Bennett - 4-072214-00 Oxygen Sensor ENVITEC Compatible O2 Cell for Covidien Puritan Bennett - 4-072214-00 Oxygen Sensor ENVITEC Compatible O2 Cell for Covidien Puritan Bennett - 4-072214-00 Oxygen Sensor

For Sale ENVITEC Compatible O2 Cell for Covidien Puritan Bennett - 4-072214-00 Oxygen Sensor

  • Asking Price :$119.95 USD [convert]
  • Condition : New
  • Weight : 1.00 Lbs
  • Qty. Available : 1
  • Date : May 16, 2025
  • Listing # : 3239106
cps-logo Get a CPS 2-Year Protection Plan starting at $5.99. Sold separately. Learn more
Advertisement

Item Description

Seller assumes all responsibility for this listing..

Listing: #3239106

  • Condition: New
  • Brand: ENVITEC
  • Type: Oxygen Sensor
  • Model: Compatible O2 Cell for Covidien Puritan Bennett - 4-072214-00
  • MPN: OOM202
THIS PRODUCT IS NO LONGER AVAILABLE.

100% New Genuine OOM202 Envitec Oxygen Sensor

MPN: OOM202

Beware of Knockoffs and Imitations from China. Our stock is guaranteed AUTHENTIC and ORIGINAL!

Ships from USA warehouse.

Features
• Compliant with European MDD (CE certification)
• Meets ISO 80601-2-55
• Designed and manufactured according to EN ISO 13485
• Accurate and reliable fast response
• Resistant to N2O
• Excellent signal stability
• High product quality
• Short delivery times
• Technical support
• Made in Germany
• FDA approved

Technical Specifications:
Measurement range 0 % ... 100 % oxygen (at atmospheric pressure)
Nominal sensor lifetime ≥ 1 000 000 % volume oxygen hours
Output in ambient air 13 mV ... 16 mV
Electrical interface 3 pin (Molex® 22-11-1031)
Accuracy meets ISO 80601-2-55 requirements
Repeatability < 1 % volume O2 at constant temperature and pressure
Linearity error < 3 % relative
Response time < 12 s to 90 % of final value
Zero offset voltage < 200 µV in 100 % nitrogen, applied for 5 min
Cross interference meets ISO 80601-2-55 requirements
Influence of humidity -0.03 % rel. per % RH at 25 °C
Pressure range 0.6 bar ... 2 bar (ppO2 0 ... 1250 mbar O2)
Influence of pressure proportional to change in oxygen partial pressure
Influence of mechanical shock < 1 % relative after a fall from 1 m
Operating temperature 0 °C ... +50 °C
Temperature compensation built-in NTC compensation
Effect of temperature compensation (steady state)
between +25 °C and +40 °C: 3 % relative error
between 0 °C and +50 °C: 8 % relative error
Operating humidity 0 % ... 99 % RH non-condensing
Long term output drift < 1 % volume oxygen per month
typically < -15 % relative over lifetime
Storage temperature -20 °C ... +50 °C
Recommended storage +5 °C … +15 °C
Recommended load ≥ 10 kOhms
Warm-up time < 30 minutes, after replacement of sensor
Weight approximately 28 grams