The BioSpec-1601 incorporates all the fundamental performances and functions required in UV-Vis spectrophotometry: Measurement at a fixed wavelength; Spectrum measurement- The wavelengths are scanned and the spectrum is recorded, which can be processed in many ways; Kinetics- Reaction rates and enzyme activity values are calculated; Quantitation- Quantitation may be performed by the single-wavelength, 2-wavelength, 3-wavelength, or derivative method.
BioSpec 1601 Specifications
*Photometric system Double-beam system
*Baseline stability 0.001 Abs. / hour or better
*Baseline flatness 0.002 Abs.
*Wavelength range 190.0 to 1100.0nm
*Wavelength accuracy 0.5 nm (Automatic wavelength correction)
*Wavelength repeatability 0.1 nm
*Wavelength drive speed About 6000 nm/min. slewing, and about 3200 to 160nm/min. scanning.
*Stray light 0.05% or less at 220.0nm and 340.0nm
*Spectral bandwidth 2 nm
*Photometric range 0.5 to 3.999 Abs. and 0.0 to 300 %T
*Photometric accuracy 0.004 Abs. at 1.0 Abs. and 0.002 Abs. at 0.5 Abs. (Tested with NIST 930D filters)
*Photometric repeatability 0.002 Abs. at 1.0 Abs. and 0.001 Abs. at 0.5 Abs
*Noise level 0.0005 A or less at 500 nm