Burdick Quest Cardiac Stress Testing System
The Quest system is right for offices looking for value, simplicity, and increased patient throughput. The system is easy to use because it features a touch screen with convenient pull-down menus for intuitive workflow. Our proprietary Logistep guide walks users through the test by highlighting the next step. We put everything you need on one screen, so there's no need for toggling among screens. And to save you time, we added an easy way to correct unintentional phase changes without having to start all over again.
Configure the system with a Cardiac Science TM55 or TM65 medical treadmill and your office will be up and running.
Features:
Eleven standard, preprogrammed protocols
Performs both resting 12-lead ECGs and stress tests
Touch screen interface
Logistep software guides testing, step by step
Option to use the University of Glasgow ECG interpretation algorithm
Compatible with Ethernet protocols, our Pyramis data management system, and our HeartCentrix connectivity solution
Warranty: 3-Year Parts and 1-Year Labor
Benefits:
Flexibility. Quest offers 11 standard, pre-programmed protocols and allows you to perform resting 12-lead ECGs and exercise stress tests.
Connectivity. The system is compatible with Ethernet protocols, our Pyramis data management system, and our HeartCentrix connectivity solution.
Accuracy. The system eliminates non-cardiac electrical noise and baseline wander for clear ECG tracings.
Simplicity. Just point and touch the intuitive touch screen – that’s all there is to it.
Ease of use. The Logistep feature walks users through the test by highlighting the next step.
Confidence. Opt for the University of Glasgow ECG interpretation algorithm to consider clinically significant criteria to interpret patient data.
Real-time. Quest's large screen instantly displays 3, 4, or 6 channels and freeze frames for detailed reviewing and printing.
Specifications:
DIMENSIONS:
60" x 25" x 34" (152 cm x 63 cm x 86 cm)
WEIGHT:
<170 lbs. (78 kg)
MONITOR
Type:
15" High-resolution color with integrated touchscreen;
Mount:
User-adjustable tilt (•15°) and swivel (•90°);
Content:
3, 4 or 6 ECG channels, real-time; information and alphanumeric data;
Displayed leads:
User-selectable during a test; (3-, 4- and 6-channel); Waveform display •8 seconds per lead of ECG in 3-channel mode
KEYBOARD
Main keyboard:
QWERTY layout;
Ergometer control:
Dedicated keys for ergometer speed, grade, start and stop
INPUT/OUTPUT
Digital I/O:
RS-232 interfaces to treadmill, bicycle ergometer, and optional automatic blood pressure monitor; TTL pulse for QRS detection, 50 msec. delay;
Analog ouput:
4 user-selectable outputs
STANDARD INTERFACES OPTIONAL
Network:
Ethernet 10baseT for full disclosure report storage and export of final report data using HL7 protocol or XML;
Removable media:
Zip drive for full-disclosure test storage – compatible with 100 MB and 250 MB disks
POWER REQUIREMENTS:
115/230V AC +/- 10%, 2.5/1.25 A, 50/60 Hz nominal 300 VA max.
CHART RECORDER
Paper type:
Pre-grided thermal paper, Z-fold, 8.5" x 11" (US letter) or 210 mm x 300 mm (A4);
Printout device:
216 mm thermal dot array;
Paper speed:
10, 25, 50 mm/sec. +/- 2%;
Printing options:
Multiple format ECG waveform and alphanumeric printing, diagnostic quality;
Gain:
5, 10, 20 mm/mV +/- 5%;
Frequency response:
0.05 Hz to 150 Hz +/- 10%;
Muscle artifact filter:
User-selectable: Off, 20 Hz or 40 Hz; Baseline filter User-selectable: Off, 0.12 Hz or STABLE (meets or exceeds all 1990 AHA recommendations);
Calibration signal:
1.0 mV for 200 msec. +/- 5%
ECG COMPUTATIONS
Heart rate computation:
8-beat average, 30 bpm to 250 bpm, updated every second
ST measurements:
20 sec. moving window with 10 sec. updates to the display and to reports;
ST parameters:
Level, slope for 12 leads simultaneously;
QRS detection channels:
II, V2 and V5 with switching;
Ectopic detection:
Ventricular arrythmias, unclassified;
ST reference:
User-selectable: E-point (QRS onset), J-Junction (QRS offset), ST point (0 msec. to 100 msec., 5 msec. interval) with display/printout annotation
PROTOCOLS
Preprogrammed:
Manual, Bruce, Modified Bruce, Balke, Naughton, Ellestad, Kattus, Ramp, Astrand, Modified Astrand, Pharmacological;
User-defined:
20 protocols with up to 30 stages each plus; Warmup and Recovery phases
PATIENT MODULE
Dimensions:
1.3" x 4.3" x 3.8" (3.3 cm x 10.92 cm x 9.65 cm);
Weight:
<1 lb. (0.5 kg);
Frequency response:
0.05 Hz to 150 Hz, +10% to -30% (maximum);
Defibrillation protection:
To 360 J, 5 kV;
Common mode rejection:
>120 dB at 60 Hz;
Electrode offset tolerance:
+/- 300 mV;
Line frequency filters:
User-selectable 50 Hz, 60 Hz or off;
Noise:
<30 μVpp;
Pacemaker pulse indication:
Per EC13;
Input channels:
Simultaneous acquisition of 8 leads plus 4 derived leads;
Defibrillation recovery:
Maximum recovery time is 8 seconds;
Patient connection:
10 replaceable leads
EXERCISE STRESS REPORTS
During exercise:
12-lead or Average Complex with Trend Graphs (HR/BP, Single-lead ST, 12-lead ST, Side by side averages, Overlaid averages, 12-lead averages), Rhythm, Ectopic Beat, Event Identification;
Final report:
User-selectable: Exercise Test Summary in tabular or narrative format; Average Complex Summary (Baseline, Maximum ST, Maximum Exercise); Trended Average Complexes (per minute or per stage); Trend Graphs (HR/BP, RPP, METS, Ectopics, 12 single-lead ST); Stress Report Form;
Comments:
Selectable, user-defined;
Configurations:
Up to 5 user-defined configurations with final reports and protocol selections
ERGOMETER INTERFACE
Treadmill:
Dedicate a 110V 20 amp plug in. Dedicated serial RS-232C for TM55 or TM65; Bike Dedicated serial RS-232C for Treadmill Burdick supported bicycle ergometer
ERGOMETER CONTROL
Control modes:
Automatic or manual; Displays Workload in METs; Phase, Stage and Exercise; Test Time (ETT); Ergometer values for treadmill-grade in percent, speed in mph or kmh; for bike -workload in watts, speed in rpm
Displays:
Workload in METs; Phase, Stage and Exercise; Test Time (ETT); Ergometer values for treadmill-grade in percent, speed in mph or kmh; for bike -workload in watts, speed in rpm
ENVIRONMENTAL SPECIFICATIONS
Operating:
Temperature 50°F to 104°F (10°C to 40°C);
Relative humidity:
15% to 95% non-condensing;
Atmospheric pressure:
525 mmHg to 795 mmHg (700 hPa to 1060 hPa)
CONFORMS TO STANDARDS:
EN60601-1-1; IEC 601-2-25; EN60601-1-2; ANSI/AAMI EC11, 1991
SAFETY:
IEC 601-1 Type CF Defibrillator Proof
WARRANTY:
Quest system, 3 years; treadmill, 13 months
This item is considered heavy/bulk and a slight adjustment may be made to the shipping fees, as per Terms and Conditions.
Shipping costs are subject to change depending on location. Pricing for international & unincorporated territories varies too based on location. Please call us for more information regarding the shipping cost for this product. Thank you!
The sale of this item may be subject to regulation by the U.S. Food and Drug Administration and state and local regulatory agencies and therefore restricts sale of this item to authorized purchasers only. Do not bid on this item unless you are an authorized purchaser. If you are unsure whether or not you are an authorized purchaser, please email us or call us. If the item is subject to FDA regulation, I will verify your status as an authorized purchaser of this item before shipping of the item.
If you have any question on this item, any other item, or general inquireis about our other cardiology specials, please email or call us at your convenience. We are open 7 days a week, 24 hours a day to respons to your needs. Our return policy is also available upon request. In short, we do whatever it takes to keep you as our customer if you are unsatisfied with your product. Credit card fees may apply.
If you do not see what you are looking for, please give us a call.
Return Policy
Items are sold as-is with no returns or refunds available unless explicitly stated.