Faulty Phenomenon
The Philips IU22 display showed blurred images with black vertical lines, persisting through multiple restarts.
Troubleshooting and Solutions
Philips IU22 Screen Flickering - Fixed
---
📺 Problem Reported
The Philips IU22 display showed blurred images with black vertical lines, persisting through multiple restarts.
---
🔧 Diagnosis & Repair Process
Systematic Troubleshooting:
1. Tested display - issue persisted with different monitor
2. Checked video conversion board - ruled out as cause
3. Swapped graphics card - problem disappeared
4. Identified swollen capacitor on original graphics card
Root Cause:
Faulty capacitor on specialized PC graphics card
Solution:
Replaced the damaged capacitor on graphics card
---
✅ Repair Result
- Clear display restored without vertical lines
- All imaging functions working properly
- System performance returned to normal
---
🔌 Technical Note
Ultrasystem PC components are often custom-made and require component-level repair when replacements aren't available.
---