William Davis - Everest Medical Solutions (HK) Co., Ltd.

Everest Medical Solutions (HK) Co., Ltd.

starstarstarstarstar speedometer75 (87)

8x
Ultrasound - ALOKA
8x
Ultrasound - PHILIPS
8x
Ultrasound - TOSHIBA

Everest Medical Solutions (HK) Co., Ltd.

Top Preferred Ultrasound Repair

William Davis, Sales Manager
DOTmed user since July 2014

G/F,West,China Fen Hin BLDG,
No.5 Cheung Yue Street
HK, Lai Chi Kok 999077 Hong Kong
Phone: +852 56190656
http://www.everest-med.com
008618664529100
welshen86
William Davis, Sales Manager
Phone:+852 56190656
May Deng, Sales
Phone:+86 18816788895

Latest Listings & Products From Everest Medical Solutions (HK) Co., Ltd.

About Us

Following are the main projects we based on:
 
Repair Ultrasound  (Ultrasound Board/ASSY Repair, Probe/Transducer Repair)
 
Repair Associated Items (Endoscopy Repair, Patient Monitor Repair, Ventilator Repair)
 
Parts & Accessories (Compatible/Original Ultrasound Transducer, Ultrasound Parts/Boards,Ultrasound Biopsy Needle Guide, Patient Monitor Module,CT part)

To support top quality repair service and most reasonable price, we have the below well-armed teams to deal with each and every service chain.
 
Professional & passionate pre-sale service team
10+ years experienced repair technician team
Reliable & responsible quality control team
Experienced & specialize after-sale technical support team

 

Besides as one distributor of high quality Medical equipment(ultrasound transducer probes,ultrasound scanner sensor,ultrasonic parts & assy, patient monitor modules,endoscope parts ,ventilator parts and other hospital related equipments),after merged to Guangzhou Rongtao Medical Technology Co.,Ltd,Everest Medical has enhanced the capacity to provide repair/refurbishing service, including repair engineers team with 10 years experiences.

Our mission is to help people realize the efficient circulation utilization of medical resources! 

We concentrate on doing two things well: First, offering a good platform to buy and sell used/refurbished/new original/new compatible medical equipments and parts, the price are competitive, and goods are tested and inspected before transaction. Second, offering fast high-quality maintenance and repair service for hospitals and other companies.

Meanwhile ,We also provide repair/refurbishing service in the following brands of ultrasound, endoscope,patient monitor, ventilator: 1.Ultrasound systems,ultrasound transducer scanner/ultrasonic probe sensor, ultrasonic board& spare part: GE, Philips, Toshiba, Hitachi Aloka, Biosound/Esaote, Philips,Medison, etc. We also provide compatible/clone ultrasound probes,biopsy needle guide, and repair service on ultrasound board/part. 2.Flexible Endoscopy and camera system: Olympus, Fujinon, Pentax, Wolf, Stryker, Storz, ACMI etc. 3.Patient monitor repair service and parts replacement: Philips, GE, Mindray, Welch Allyn. We provide LCD screen, power supply, PCB board of the patient monitor.4.Ventilator air module and oxygen module repair.

We have rich stocks,10,000+pcs original/compatible clone new ultrasound transducers, ultrasonic boards, ultrasonic biopsy needle guide,CT/LUNAR/RAD/MAMMO Parts,patient monitor boards and modules.

 

Our aim is to offer cost-effective and good-quality medical equipments and spare parts, as well as repair service in ultrasound ,patient monitor endoscope,ventilator and other related medical fields.

Our Vision is to be the top preferred third-party service company worldwide ,thank you for your continuous support!

Equipment Categories

Equipment

Bedside Monitor
Bronchoscope
Cardiac - Vascular
Co2 Monitor
Colonoscope
Cystoscope
Defibrillator
Endoscope
Endoscopy Processor
Gastroscope
Module
Monitor
OB / GYN - Vascular
O/R Camera
Oxygen Sensor
Scope Accessories
Shared Service
Software
Ultrasound Accessories
Ultrasound General
Ultrasound Transducer
Vascular - Small Parts
Ventilator
Vet. Ultrasound
Video Endoscopy

Services

Distributor
Parts, New, Used and Rebuilt
Pre-Owned Equip Remarketer
Repair, Maintenance, Service

Purchases

All the medical devices & parts are in “As is “condition, which will be declared clearly in the purchase order.

Payment

100% prepaid before shipping,we accept bank transferring and western union.

Delivery

Worldwide shipping.EMS/DHL/Fedex Express will be arranged in 1-5 working days unless otherwise specified.We Seller is responsible for export Customs clearance in our end,and buyers for import Customs clearance in their end.

Returns

Goods are permitted to return in the same appearance and operation condition within 7 days. Returned items are subject to 25% restocking fee.Shipping fee should be borne by purchasers. No returns after 3 days for goods that had been ordered wrong.

How to keep your ultrasound transducer well?

To make sure your ultrasound transducers/probes are in good keep, you should pay attention to the following matters:

  • Temperature (suitable)
  • Humidity (suitable)
  • Environment (cloth, paper, human body)
  • Coupler (mind quality, avoid corrosion)
  • Accidental damage (avoid damage from mouse and cockroaches)

Five Steps to repair & refurbish your ultrasonic board/part

1.Check before sending to us

  Make sure your defective/broken ultrasound board/part is complete and never repaired.

2.Delivery

  Send it to our appointed address for receiving package from abroad after well-wrapped.

3.All-round inspection

  Our repair engineers will do all-round inspection once the ultrasound board/part reaches, then provide suitable repair solution.

4.Repair and Refurbishing

  For general/minor troubleshooting, 1-3 days to fix. For major/difficult repair, 4-7 days to fix.

5.Send the fixed board/part back

  After strict testing before delivery, the fixed board/part will be sent back in good working condition with 60 days warranty.

How to do complete clinical evaluation of your transducer?

1.Frequency conversion

 Whether the frequency can be&&&& transformed,it:s&&&& the important indicator.

2.Frequency range

 The Wider the range,the better.

3.Scanning angle

 The bigger the angle, the better. Especially for the intracavitary probes.

How to clean air filter of your diagnostic unit?

Ultrasound diagnostic unit maintaining is not a simple thing,today I'm going to share you some ideas of cleaning the inside air filter.

First, cut off the power, then dismantle the air filter. follows are several matters need your attention.

1.Before installing the air filter, make sure it is dry. Humid one may cause trouble of the whole diagnosis system.

2.Don't open the power supply at the same time.

3.Remove the dust with blower,or use a moist cloth.Dry it after finished.

Notes: Usually we should clean the air filter per one year,but in high-dusty room,the suggested interval time is every 90-180 days.

Notices for keeping Ultrasound Device

1.Power

The voltage should be within ±10 volts,choose voltage-stabilized source,independent power only.

2.Prevent short circuit

Cleaning the device regularly,prevent short circuit.

3.Replace transducer

We should freeze or shut down the device before replacing transducer.

Repair Philips HD 15 Analog Interface Moudule

System: Philips HD15
Part name: Analog Interface Module (AIM)
Part number: 453561197306

Failure phenomenons:
Error occurred after started up, customer confirmed AIM board was broken down.

Solutions:
Repair and replace the AIM board

 Common faults of GE ultrasound machine

System

Faults

Solutions

LOGIQ  500

No image when scanning far field

Check power supply, reinsert the circuit board. Suggest to do metal earthing.

 

LOGIQ 500PRO

Artifacts appear in color flow imaging (betweentimes)

 

Replace ACFP3 board

LOGIQ 200E

No image on displayer, whereas normal image in workstation

Fix displayer

LOGIQ 200E

Curved lines interpolated into sector-shaped image, from near field to far field

Replace DSC board, and make sure the system earthing

LOGIQ 200

No sound for the probe, requires multiple startup

Detect low voltage and external power supply (MST)

LOGIQ BOOK

Lithium battery- short life

Replace lithium battery, and upgrade the system to 2349040-23 KB824146

LOGIQ BOOK

Crash


Reinstall the system

LOGIQ 400MD

cardiac probe shows clutter filte, interference in far field

Repair cardiac probe S317, then test the voltage

LOGIQ 400MD

Noisy in probe, no waves in scan area

Connecting board failure, try to connectwith TRDR and HBFR

LOGIQ 400CL

Automatic interruption

High voltage failure, need repair

LOGIQ 180

Screen wobble

Displayer failure, need stabilized voltage supply

LOGIQ 500MD

No image after boot (betweentimes)

Detect five low voltage value

LOGIQ 200PRO

No image

Detect the channel board

LOGIQ 9

No scan area, no ultrasound image

MSTE and high voltage power supply failure, need repair

SIEMENS S1000 system maintenance

Failure phenomenon:
Siemens S1000 color super machine, the boot into the ultrasonic working interface is normal, the heart probe is used normally, but the superficial probe and the abdominal probe can only be used normally in the 0B mode, can not be used in other modes, the machine has a card screen phenomenon at the same time.

Failure analysis:
The signal processing and control module of S1000 color supercomputer is mainly composed of probe interface module, input and output module and electronic system module. The electronic system module includes TR, CB, RC, BE, VI and RM board, TR, RC board and probe interface module ( The TI Module) performs the ultrasonic signal processing functions together, collectively referred to as the front end portion; the BE, VI, and RM boards are responsible for the subsequent processing of the ultrasonic signal data and the overall control functions of the machine, collectively referred to as the back end portion. The power supply consists of four sections, an AC AC power box, a PSA analgo power module, a PSD digital power module, and a QuickStart Battery.

Firstly, if the system software error occurs, the fault occurs. After the system is reinstalled, the fault still exists, and the system software fault may be eliminated. The problem of RC and BE is initially judged from the hardware function. As the real-time control board of the front end, the RC board is also responsible for the control function of ultrasonic transmission and reception. It is also responsible for analgo-to-digital conversion of the 192 channels of received data of the TR board, thereby converting the echo signals into digital signals, which is convenient for the back-end software. Further processing of the technology. A failure of the RC board will cause all other functions of the machine to be abnormal. The BE board receives the digitized echo signals from the front RC board and processes them according to the operator's requirements. The failure of the BE board often causes one of the functions of the color Doppler to be abnormal.

The PSA and TI can be considered by the hardware replacement exclusion method to eliminate the BE and RC failures. The TI board can support 3 probes at the same time and can be switched at will, but the probe hot swap is not supported. When the TI board fails, it will cause the 2D area to have no image, card screen or boot code "US_IMG_17" and other error codes. The PSA is responsible for the front end power supply and provides the probe working voltage. If the PSA fault will cause the probe to be caused by the abnormal voltage supply. The function is not working properly.


Repair results:
Replace the PSA power supply, all the functions of the machine are used normally, and the fault is eliminated.

GE VIVID I system maintainance

First, the failure phenomenon:
The machine starts normally, and after receiving the probe, it prompts an error and the probe is not recognized.

Second, the failure analysis:
The ultrasonic probe does not recognize the power supply that normally considers the probe interface board or supplies power to the probe interface board. The machine configuration diagram shows that there is no high-voltage switch on the interface board of the model, that is, it does not affect the high-voltage output of the DC power supply, so it can pass the detection. The DC output voltage directly determines whether there is a fault in the power supply. After the detection, the machine is found to have no high-voltage output, and it is determined that the DC power supply is faulty.

Third, testing and repair:
Replace the DC-DC power supply, the probe can be recognized normally after the machine is started, and each function resumes normal operation.

In genneral, we are really experted in ultrasound repair.

Aloka Alpha 10 system Maintenance

Fault conditions

The ALOKA &α10 machine abdomen probe has no echo, the superficial probe has only a weak echo, and the heart probe has a normal echo.

Failure analysis

Since the heart probe has a normal echo, it is necessary to first eliminate the fault caused by the probe itself, and replace the other abdominal probe for testing. The test result also has no echo, and the fault is determined by the hardware of the machine itself. There is a voltage output indicator on the IO board at the rear of the machine. It is found that the +70V indicator is not lit after the machine is started, and the +90V indicator is always on. +70V, +90V supply the working voltage of the probe to the machine, combined with the echo of the abdomen probe and the normal echo of the heart probe. The fault may be caused by insufficient voltage supply of the high voltage power supply.

Troubleshooting

After replacing the high voltage power supply EU-6031, the machine resumes normal use.

 

Repair Philips CX50 Blue Screen Fault

First, the failure phenomenon:
The machine has a blue screen when it is turned on and cannot be started normally.

Second, the failure analysis:
On-site consultation using the doctor learned that the machine did not make any changes to the machine hardware and software before the blue screen appeared, the blue screen code is 0x000000ED, the code generally points to the hard disk failure. After the machine is disassembled, the hard disk of the machine is detected. It is found that the hard disk has bad sectors, and the bad sectors of the hard disk often cause the hard disk to work normally and affect the startup of the machine. Since the direct repair of bad sectors of the hard disk will result in the loss of data of some sectors of the bad track, an attempt is made to check and repair the data of the hard disk to ensure that the color system software carried in the hard disk can operate normally.

Finally, testing and repair:
Repair the system hard disk and work normally after the machine restart.

Phililps CX30 System Maintenance

The phenomenon of failure:
The system is properly starting-up, but it will crash during use.

Failure analysis:
Crashing is mainly failure of main board, and the stuck interface prompts the discovery of a new keyboard. This may be a system driver failure resulting in incompatibility or a short circuit in the numeric keypad.
Reinstall the system, the fault remains the same, consider the numeric keypad failure, but the same problem occurs when disconnecting it from the control panel. This fault occurs only when users uses panel keys,then it is suspected that there is a short circuit in the control panel part. So it needs further testing.
The ultrasound machinevuses Windows XP system, then using a keyboard test software of the system can accurately detect whether the keyboard has any problems.
Even if the the numeric keypad and the control panel is disconnected, the failure is the same. So it can be determined that the fault is caused by the short circuit of the panel, which continuously triggers the system crash caused by the numeric keypad.

Testing and repair
Replace the control panel and the machine will return to normal.

Philips HD15 Anolog Interface Module

System: HD15
Part name:Analog Interface Module (AIM)
Part number:453561197306

Failure phenomenons:
Error occurred after started up, customer confirmed AIM board was broken down.

Solutions:
Repair and replace the AIM board

Aloka Alpha 7 RX Board Diagnosis

System: Aloka Alpha 7
Part number:EP539100BB, EP539501DG, EP539500DE

Failure phenomenons:
Few lightspots occurred on the image

Solutions:
Diagnose from EP539100BB, EP539501DG, EP539500DE

SIEMENS S1000 system maintenance

Failure phenomenon:
Siemens S1000 color super machine, the boot into the ultrasonic working interface is normal, the heart probe is used normally, but the superficial probe and the abdominal probe can only be used normally in the 0B mode, can not be used in other modes, the machine has a card screen phenomenon at the same time.

Failure analysis:
The signal processing and control module of S1000 color supercomputer is mainly composed of probe interface module, input and output module and electronic system module. The electronic system module includes TR, CB, RC, BE, VI and RM board, TR, RC board and probe interface module ( The TI Module) performs the ultrasonic signal processing functions together, collectively referred to as the front end portion; the BE, VI, and RM boards are responsible for the subsequent processing of the ultrasonic signal data and the overall control functions of the machine, collectively referred to as the back end portion. The power supply consists of four sections, an AC AC power box, a PSA analgo power module, a PSD digital power module, and a QuickStart Battery.

Repair results:
Replace the PSA power supply, all the functions of the machine are used normally, and the fault is eliminated.

Toshiba SSA-660A RX (PM30-32733) board maintenance

Date: March 3, 2015
Brand: Toshiba SSA-660A Model: RX (PM30-32733)
Symptom: Toshiba SSA-660A color ultrasound boot image is missing
Service content: Providing Toshiba SSA-660A color ultrasound RX (PM30-32733) board to judge if there is something wrong with the board
Service period: 1 working day

Toshiba SSA-580A Keyboard(BSM31-3078) Maintenance

Brand: Toshiba SSA-580

Model: BSM31-3078

Symptom: Some buttons on the Toshiba SSA-580A numeric keypad are out of order

Service Content: Repair and Replace Toshiba SSA-580A New Digital Keyboard (BSM31-3078)

Service period: 5 working days

ALOKA SSD-3500 Fault Repair

1.The fault phenomenon

After closing the switch, ssd-3500 color ultrasound was started. After the "Aloka" sign appeared in the startup process, the blue bar kept moving and the panel light kept shining. The system could not enter the normal ultrasonic diagnosis interface.

2.Failure analysis

When this happens, the first thing to think about is the CPU and HDD issues that are directly related to system startup.Since the machine's data processing platform is based on Microsoft's Windows operating system, just like ordinary computers, the boot process is basically the same.
In addition to the normal CPU and hard disk data reading in the startup process of the ultrasonic device, each image processing board containing flash should be confirmed for data, so as to determine whether there is any data loss or need to update data, so as to brush flash again in the startup process.The indicator light of the hard disk is flashing constantly during the boot process, indicating that the data of the hard disk is read and written constantly and the hard disk is troubleshooting.
Next, unplug and unplug each group of wires. After re-plugging and unplugging, the fault will remain after starting up.If it is suspected to be a CPU fault, take out the CPU board and observe it to see whether there is obvious burning down or insufficient heat dissipation causing component damage. After observation, it is normal.
After the discovery of the motherboard has a button battery, removed from the measurement of only 1.6v.It is inconsistent with the actual mark of 3V, and the voltage value is obviously not enough to support the normal reading and writing of the hardware part of the chip, resulting in problems in the data communication between the CPU and the hard disk, affecting the normal loading and data update of the device.

3.Exclusion and summary

After replacing a standard 3V battery, the device starts normally and enters the ultrasonic diagnosis interface for troubleshooting.
For ssd-3500 color doppler ultrasound startup, the blue bar after the "Aloka" sign& appears in the startup process keeps circulating and the panel light circulates and shines. The system cannot enter the normal ultrasonic diagnosis interface, which basically determines that the components or hard disk on the main board have problems.

ALOKA α5 color doppler ultrasound fault repair case

ALOKA alpha 5 is equipped with advanced front-end technology, original data processing platform,high-capacity digital beam generator (12bit digital beam generator) and other technologies to ensure the high-quality image of all organs in the body.And as well as the varties probes in ALOKA, ALOKA alpha 5 can meet the needs of linical diagnosis and scientific research,At present, ALOKA alpha 5 is used in many hospitals in China. Now we will analyze one of the common faults encountered in the use of ALOKA α5.

Faults phenomenon

When operating the CW(continuous doppler), the diagnostic instrument crashed with an error prompt, and the other functions were all normal.Error warning: the machine detects abnormal voltage, please restart the system.

Failure analysis

(1)The CW function of the diagnostic instrument can only be used in cardiac, and it has no such function in other conditions, so the CW board may be damaged.
(2) in case the appearance of the fault, the working voltage of CW measured by the attached power box drops significantly, which may be caused by the attached power box.
(3) this fault may also be caused by CW signal and loss of communication function.
(4) the heart probe (phased array probe) is composed of PW and CW, and the heart probe may be damaged.
(5) the software system is damaged, resulting in the loss of CW function.

troubleshooting

It’s not working after replacing the attached power box and related circuit boards,the dead phenomenon didn’t appeared when the heart probe was replaced . The problem was e circuits of the machine appeared,the restart of the machine are required.

GE Logiq P6 color ultrasonic fault

1. Faults phenomenon

After reading the tags, you will not be able to enter the operation interface.

2. Faultsanalysis and troubleshooting

The loss of files is usually a system fault, and the system can be restored after reinstallation. According to the customer's feedback, there are important patient information and parameters in the machine. In order to maximize the customer's interests, we try to repair the system.
Since the prompt file is missing, it is theoretically possible to recover the lost or damaged file by filling in the missing or damaged file. At the same time, because the user cannot enter the operation interface normally, the abnormal information should be found through the log file.The operation steps are as follows:

Step 1: according to the prompt path, it can be known that this file is the user's default file. After opening this file, it is known that only some basic information and Settings of the machine are recorded in it, and relevant parameters required for diagnosis are not affected.

Step 2: copy the same file in another system for replacement, connect the machine to start up and use normally, and help the customer to backup the data again.

3.Summary

Color super data should develop the habit of regular backup, so as not to cause unnecessary trouble when failure.

Maintenance of GE LOGIQ P6 Color Doppler

Fault phenomenon

All probes of GE LOGIQ P6 machine have interference in color Doppler mode.
 
Fault analysis
LOGIQ P6 machine plate structure is simple, plate function integration is high. The color Doppler processing function of the machine is on the SYSCOM motherboard. Generally, this kind of interference is caused by probe failure, external interference or SYSCOM motherboard and APS power supply.
Due to the interference of all probes in the machine, the possibility of failure of all probes is very small, so other probabilities should be considered first. Pushing the machine to another floor room for testing, there are still interference faults to eliminate the possibility of external interference. APS power supply failure may cause interference of machine access voltage. APS power supply failure may be eliminated after replacing the test. Finally, the SYSCOM motherboard was replaced, and the machine started to work normally. All the probes entered the working mode of color Doppler without any interference and troubleshooting.

Summary
GE LOGIQ P6 is easier to maintain than other ultrasonic machines because of its simple plate structure. Maintenance is mainly based on the specific performance of the machine fault and the working principle of the machine to analyze the fault location and take corresponding treatment measures, for the more difficult to analyze and judge the fault location should be carefully and step by step investigation, avoid random attempted maintenance

GE Logiq P9 maintenance

 

Fault phenomenon
Dark lines appeared in the ultrasound images. There is a dark line on the ultrasound image of GE Logiq9 color Doppler Ultrasound. There is no echo signal in the dark line area. When the image is formed, the position of the probe is moved, while the position of the dark line is always fixed.


 Fault analysis
After scanning the machine probe with the coupling agent, it is found that the edge of the dark line of the
image is obvious. According to the analysis of the principle of the formation of the ultrasound image and the structure of the color Doppler ultrasound, the cause of the failure may be the problem of the corresponding probe wafer or the corresponding channel. Switching different probes for testing, it is found that each probe has the same dark line, and each probe has the possibility of wafer damage at the same time. It can eliminate the damage of the probe, and the failure may occur on the corresponding channel board. There are 8 channel boards in the front end of GE Logiq9 color Doppler Ultrasound. The corresponding ultrasonic transmitting and receiving and A/D conversion are carried out. Eight channel boards correspond to different probe channels. Each board processes 24 channels. Mark the corresponding dark line position, shutdown and switch channel board position on the display image. When switching to the fourth channel board, the corresponding dark line position on the image moves. Then cross-check with another channel board and the fourth channel board to determine that the fault occurs on the fourth channel board.
 
Repairing
The signal processing channel boards of modern high-end color Doppler Ultrasound usually have 2, 4 or even 8 boards. These channel boards are identical. When the image is dark, the positions between the channel boards can be changed. The replacement method is used to achieve the purpose of diagnosis and lock the fault parts accurately. The machine returned to normal after replacing the fourth passage board.

Everest Medical Solutions (HK) Co., Ltd. Gold Parts Vendor Mini Profile

We are persistently striving to offer cost-effective and good-quality repair service & outright selling on ultrasound transducer probe/part/ASSY,patient monitor module,endoscope part,CR/RAD,etc.

Standard Warranty Terms:
60 days warranty unless otherwise specified

Standard Return Policy:
15 days after goods received

Years In Business: 6

Percentage Of Revenue That Parts Account For: 30%

Restocking Fee: 15%

Warehouse Size in Sq. Ft.: 12,500

Number of Parts Stocked: 15,000

Specialties

  1. Original/clone/compatible Ultrasound transducer probe,ultrasound parts/board
  2. ultrasound biopsy needle guide
  3. patient monitor device module/parts/accessories
  4. ventilator Oxygen module, air/luft module,boards
  5. flexible endoscope parts,O/R camera head
  6. new/used/refurbished original/compatible clone ultraosund transducer probe
  7. ultrasonic boards/parts/ASSY repair,exchange and for sale
  8. CT/LUNAR/RAD/Mammo Part/ASSY

Everest Medical Solutions (HK) Co., Ltd. Gold Parts Vendor Testimonials

From a dealer

Feb.25,2017

great/easy to make bussiness with
From a doctor

Apr.10,2018

This friend is surely reliable and trustworty

By an engineer

Jun.8,2018

Nice company to work with,great experience
From a dealer

July,28,2018

Nice repair work,highly recommended service company
From a hospital

Oct.15,2018

Excellent repair skills,bring our Aloka F75 Ultrasound back to life!
From an associate in Asia

Oct.26,2018

Professional aftersale service,pretty nice Philips CX50/EPIQ probes
From a dealer in Europe

Jan.10,2019

100% trustworthy,nice to work together on GE/Aloka/Hitachi Ultrasonic board repair
From a partner

May,9th,2019

Perfect repair skills on Aloka Alpha boards, EP539100/EP5555---
From a technician in Germany

Oct.12th,2019

The power and board for GE received works good.
Posted 446 Days Ago
William Davis Posted to the Service Technicians forum:
re: re: Philips CX50 service Manual
Hi,Moussa,... Click here to read more
Posted 446 Days Ago
William Davis Posted to the Service Technicians forum:
re: re: Prosound Alpha 5 service manual
Hi,Dmitry,... Click here to read more
Posted 623 Days Ago
New arrival---Used Hitachi EUP-S50A Phased Ultrasound Transducer
Posted 623 Days Ago
New arrival---Used Hitachi EUP-S50A Phased Ultrasound Transducer
Posted 738 Days Ago
ALOKA EP556600AB Ultrasound General
Just posted a ALOKA EP556600AB Ultrasound General on DOTmed... Click here to read more
Posted 738 Days Ago
ALOKA EP495000HH Ultrasound General
Just posted a ALOKA EP495000HH Ultrasound General on DOTmed... Click here to read more
Posted 1234 Days Ago
Weekly Special

Used Philips C5-2 Convex Probe for HD11/Envisor Outright price:$1,450

Used Philips L12-3 Linear Probe for HD7/HD11
Outright price:$1,560
Posted 1408 Days Ago
William Davis Posted to the Honest & Dishonest Dealings forum:
re: XMedical references
We ever had business with Valentina in 2014,seems it's ok.Hope you didn't meet problems with XMedical.... Click here to read more
[x]
Chat with the Owner

We're Offline

Send message to seller