Anton Spaci - Freelance - DOTmed Profile
advertisement
Current Location:
>
> Anton Spaci


Anton Spaci - Profile
TSHP
Freelance
Germany
DOTmed User since June 2012
Email Anton

Rate This User
(88) (33)
Anton Spaci Germany TSHP

Anton Spaci - Freelance

I work for my own with medical equipment I can provide advise and parts for toshiba equipment Can also give advise for charity in poor countries

Overall Feedback Received

No ratings.

Recent Forum Posts

XC error in Acquilion 16
November 13, 2014 04:42

check gommu plate between SRUH and tube and after make block test ...Click here to read more

No Massage Error is coming in toshiba while starting scan
November 13, 2014 04:42

"no message" means no connection between gantry and reconstruction box . As usual could be GCIF board in gantry or DTB board in reconstruction box put laptop on rtm or check in log file BIR ...Click here to read more

Asteion S4
November 13, 2014 04:42

connect laptop to your rtm and have a look maybe bootlines are not ok ...Click here to read more

  All forum posts by Anton Spaci