VENDOR DESCRIPTION
The IMED Gemini PC-1 is a single channel volumetric infusion pump and controller which provides accurate and automatic infusion of intravenous drugs and fluids.
FEATURES
Backlit center display
Rapid Rate, On-Line Titration
Infusion Rate Profiling Options
Panel Lock
Pressure History
Drug Dose Calculator
Programmable Start Time
Multi Dose
Volume/Time Dosing
SPECIFICATIONS
Operating Principle: Linear peristalsis
Dimensions:- 10.8" x 8.2" x 6.5" (27.5 cm x 20.6 cm x 16.5 cm)
Weight:- 11.3 lbs (5.1 kg)
Power Requirements:- 90 - 130 (120 nominal) VA, 0.5A, 50/60 Hz, 3 wire, single phase
Battery Operation: With a new battery, approximately 5 hours At 125 ml/hr before a "HELP BATTERY"
Electrical Leakage:- Less than 100 micro amps
Delivery Rate Range
Micro:- 0.1-999 ml/hr (Note: Recommended maximum rate in the Controller Mode is 500 ml/hr
Macro:- 0.1-99.9 ml/hr (in 0.1 ml/hr increments)
Volume to be Infused
Micro:- 0.1-999.9 m/l (in 0.1 ml increments)
Macro:-1-9999 ml (in 1 ml increments)
KVO (Keep Vein Open) Rate:-1 ml/hr if set rate is 1 ml/hr or above, or actual rate if set at 0.9 ml/hr or below
Occlusion Pressure
Pump mode:- 10 ±2 psi for rates above 30 ml/hr. For rates of 30 ml/hr and below, the occlusion pressure will be less than 10 psi, and is rate dependent (unless locked into 10 psi for all rates)
Advisories:- Secondary, Primary, Infusion Complete - KVO, Empty Container-KVO, Low Battery, Maintenance, Pause, On Batt, 10 psi. Flow, Total Vol Infused, Sec Vol Infused Macro, Micro, Versataper: Setup, Step: xx, Steps: xx, Total Vol: xx, Duration: xx, Versataper, and Time Left: xx
Prompts: Select P/C mode, set rate, set VTBI, and Press Start
Alarms: Check ECD, Check IV Set, Close Door, Air-in-Line, Occluded Patient Side, Occluded Fluid Side, Occluded, Partial Occlusion Fluid Side, Flo-Stop Open/Close Door, KVO Comm Cable, KVO Comm Tout, and KVO
Malfunctions: Help internal error, help battery, and (Watchdog Malfunction)
Secondary (Piggyback): Dual rate programming