Fluke VT Plus HF Gas Flow Analyzer in good overall condition. Unit powers on normally, display is clear, and controls respond properly. Functional testing has been performed to verify operation.
During operation, the unit displays a Linearization Table Invalid error message. The message text is partially truncated on screen and full details cannot be accessed via the interface.
Despite this warning, the analyzer demonstrates active flow and pressure measurement during testing (see testing details below).
Testing PerformedPower-On Test
Unit boots successfully.
Performed with no external device connected to evaluate internal pneumatic integrity.
Results:
Stable pressure reading
Airflow introduced manually through tubing.
Results:
Analyzer detects airflow events (breaths)
Unit displays:
Linearization Table Invalid error
This indicates internal calibration lookup tables may be missing or corrupted. The analyzer still detects airflow and calculates volume during testing; however, measurement accuracy relative to factory calibration is unknown.
Unit may require calibration software/service to restore linearization tables for full specification accuracy.
IncludedFluke VT Plus HF Gas Flow Analyzer
Calibration status unknown.