Get a CPS 2-Year Protection Plan starting at $5.99. Sold separately. Learn more
Seller assumes all responsibility for this listing..
Listing: #5643501
***Listing includes everything that is pictured unlessotherwise specified***
This item has been biomed tested and verified to be ingood
physical and working condition. 30 Day Warranty.
POWERS ON, DISPLAY- PASS, BUTTONS- PASS, TONE- PASS, SMOKE EVACUATOR POWERS ON. FUNCTIONAL CONTINUOUS OUTPUT TEST DONE ON RIGEL UNITHERM: OUTPUT LOAD SET TO 500 ohm; WITH COAG POWER SET TO 80W=440mA**OVER 400mA(+5%): FAIL**; PURE CUT SET @ 10W = 148mA; BLEND SET @ 100W = 470mA. OVERALL TEST IS A FAIL, OUTPUT RANGE IS HIGHER THAN SPECS. UNABLE TO FURTHER TEST. RECOMMEND CALIBRATING. CONDITION: USED, AS IS. INCLUDES FOOTSWITCH.
Please look at pictures carefullybefore purchasing.