Baxter 6201 infusion pump error code F_82
August 17, 2016 10:44
"No acknowledgment message from Slave CPU for three communication trials." in Service Manual. checks : Turn device off and back on. Corrective action ; If failure recurs, Replace CPU board....Click here to read more