Problem in GE AMX4 Plus, 45E Generator Error
September 15, 2022 09:56
Try these fixes. ERROR 45E 1. Check For 2 KHZ Phase 1 & 2 Signals At J2-17 & 19 Of FIL/KVP Board 2. Check For Continuity Between J5 Plug On FIL/KVP Board And Filament Terminals On HV Xformer 3. Check For Open Or Shorted Tube...Click here to read more
GE AMX 4 Error Code 102 - Charging
June 09, 2022 11:40
Hello Gary, I would first check your batteries to be sure one of them is not bad. Next, I would try a recalibration of the charging circuit. Finally, if those do not solve your issue replace the charging board. Hope this helps,...Click here to read more
GE AMX4 Service manual
April 30, 2021 09:54
You can find that manual on the GE documentation site. https://customer-doc.cloud.gehealthcare.com/#/cdp/dashboard...Click here to read more
G E AMX4 Generator Fault.
December 30, 2020 10:56
Hello, below are the errors you asked about. Unfortunately, I do not have much on the 106 error ERROR 45C 1. Check For 80 VAC At CR4 Black Leads On AMX1A4 Panel With PREP Pressed 2. Check For About 75 VDC Across AMX1A4 C...Click here to read more
GE Amx 700 charging level.
May 13, 2020 09:27
First, did you replace the batteries? After two years sitting and not being used I am sure there are a few that has gone bad. Once that is completed. You should be good to place it in charge mode. However, you have to let the mac...Click here to read more
GE Innova 4100 Service Manual
February 28, 2020 10:20
Try the GE Document library https://www.gehealthcare.com/products/support-documentation-library ...Click here to read more
GE AMX 4 Plus Portable X-Ray Machine
February 21, 2020 10:35
During Calibration: 1. Check If K7 Contactor On AMX1A4 Panel Pulls In - Check For Welded Contacts 2. Check That No Wires Or Side Cover Insulation Interfere With K1 Thru K6 Contactors On AMX1A4 Panel 3. Check For Bad Crimp On R...Click here to read more
GE Optima XR 220 User and Service Manual
May 08, 2019 10:22
Good morning, You may want to try the GE document library they have the white CD information on that site and should be able to cover most problems. https://www.gehealthcare.com/en/support/support-documentation-library...Click here to read more
GE Advantx error 41e1 on exposure
January 30, 2019 10:01
Good morning. give this a look. MESSAGES Console Error Message: Power unit control not operable. Call service. InâRoom Display Message: POWER UNIT NOT OPERABLE MODULE SCPU type power unit DESCRIPTION This error is ge...Click here to read more
GE LOGIQ e: Ghost Image and App SW 5.0.1 needed
June 03, 2013 11:08
Also, may I suggest that you obtain a disk duplicator and on the next pm you pull the hard drive and duplicate it in case this happens again. This is a good idea for all systems as a quick backup repair. furthermore, put the dupli...Click here to read more