Over 150 Missouri Auctions End Tomorrow 06/05 - Bid Now


Omar Ortiz - Profile
Regional Service Manager
Medimax
PR, Puerto Rico
DOTmed User since October 2011
Email Omar

(36) (14)

Overall Feedback Received

No ratings.

Recent Forum Posts

SES error on my Bennett 300 SE
January 31, 2014 04:24

SES mean Store Energy System, you may have problems with a fuse blown on the charging system a bad power supply, a bad cables connector on the batteries or maybe a bad battery bank. At the battery section their are some fuses, tak...Click here to read more

Lorad M-IV hologic mammogram showing "System Error 35"
October 29, 2013 11:08

this means tube current is below 10% of the expected value. Basically mA problem. It happens in Small and large filaments? You need to check, filament cal, filament board, power supplies to the filament circuit, etc... ...Click here to read more

Imag/Study Transfer to View Forum Computer
October 25, 2013 07:34

hi, I don't know this unit but something similar happens to me in te past with other digital unit. if this is a Digital receptor system maybe the wake up time for the detector in the mornings is wrong and the detector is not ready...Click here to read more