Image/Ring artifacts
There are several conditions that can cause CT image/ring artifacts. More often than not, artifacts are indicative of a DAS (Data Acquisition System) malfunction, not a problem with the tube. One quick and easy solution is to check the tube alignment. Beam cut-off or lower tube output occurs when a tube is out of specification. Refer to the OEM guidelines for proper tube installation and follow the alignment procedures.

Sign-up for Dunlee's free E-News Bulletin and join the inner circle to receive information on new replacement tubes, upgrades, useful technical tips, and relevant industry news.