by
Barbara Kram, Editor | February 14, 2007
Recalled infusion pumps
were formerly known as Alaris SE
The U.S. Food & Drug Administration (FDA) has announced that Cardinal Health 303 Inc. (Cardinal 303), formerly known as Alaris Medical Systems, Inc., and three of its top executives, have signed a consent decree for condemnation and permanent injunction related to their Signature Edition (SE) infusion pumps. The infusion pumps have a design defect referred to as "key bounce" which may cause the pump to recognize a single key stroke as a double key stroke. The "key bounce" problem poses a risk to public health due to a potential over-infusion of medications.
Cardinal 303 has agreed to stop manufacturing and distributing all models of the SE infusion pumps until Cardinal 303 corrects manufacturing deficiencies and until the devices are made in compliance with the current good manufacturing practice (CGMP) requirements and the Quality System (QS) regulation for medical devices.
Infusion pumps are electronic devices intended to control delivery of solutions and medications to patients. They are used in situations where medication must be administered intravenously or through other routes, in a continuous or intermittent manner, for a prolonged period of time.

Ad Statistics
Times Displayed: 109400
Times Visited: 6640 MIT labs, experts in Multi-Vendor component level repair of: MRI Coils, RF amplifiers, Gradient Amplifiers Contrast Media Injectors. System repairs, sub-assembly repairs, component level repairs, refurbish/calibrate. info@mitlabsusa.com/+1 (305) 470-8013
Under the terms of the consent decree, the company has agreed to take necessary measures to ensure compliance with the CGMP requirements and the QS regulation by all of its facilities that design, manufacture, process, pack, label, hold, or distribute SE infusion pumps. The decree was signed by Dwight Winstead, Cardinal 303's President and Chief Operating Officer, David L. Schlotterbeck, the company's Chairman and Chief Executive Officer, and William H. Murphy, the company's Senior Vice President of Quality and Regulatory Affairs. The decree also requires the company to retain an independent expert consultant to conduct inspections of its SE infusion pump facilities and certify to FDA that corrections have been made. FDA will continue to monitor these activities through its own inspections.
Under the consent decree, FDA will allow Cardinal 303 to continue to service and repair SE infusion pumps that were already in the hands of customers before entry of the decree. The company is also required to submit to FDA an acceptable detailed corrective action plan to bring the SE infusion pumps currently in use in the United States into compliance with the Federal Food, Drug, and Cosmetic Act.
The decree was entered in the United States District Court for the Southern District of California on February 8, 2007.