Barbara Kram, Editor | March 26, 2007
ARLINGTON, VA- The Association for the Advancement of Medical Instrumentation (AAMI) has released the first-ever standard that specifies requirements for container systems and containment devices that hold sterilized surgical instruments, building on a record number of new and revised sterilization standards and guidance documents that have been released or published over the last year.
The new standard-ANSI/AAMI ST77:2006-Containment devices for reusable medical device sterilization-provides requirements to help ensure that containment devices allow sterilization and sterility maintenance of their contents, as well as (where applicable) adequate drying and sterilant removal.
Nancy Chobin, RN, of St. Barnabas Health Care System in West Orange, NJ, and a member of AAMI's Reusable Sterilization Container Working Group, calls ST77 "a long-awaited and sorely needed document," because there was previously "no regulation or guidance on the manufacture, cleaning, or sterilization of rigid containers or organizing cases...Both the materials and the construction of the containers and organizing cases has an impact on sterility maintenance because the design often would cause damage to the packaging system being used."

Ad Statistics
Times Displayed: 19090
Times Visited: 362 Stay up to date with the latest training to fix, troubleshoot, and maintain your critical care devices. GE HealthCare offers multiple training formats to empower teams and expand knowledge, saving you time and money
ST77 addresses these important concerns about materials of construction by setting requirements for a number of key parameters, including:
* Durability-materials shall not deteriorate (crack, flake, peel, fracture, become brittle, or deform) within the manufacturer's recommended useful life at the maximum conditions of use.
* Compatibility with the sterilization process-materials shall not inhibit or interfere with the sterilization and drying process for which the containment device is recommended.
* Biocompatibility-materials shall not adversely affect the biocompatibility of the devices being processed.
The list price for ANSI/AAMI ST77:2006 is $90; the AAMI member discount price is $45. To order, call (877) 249-8226 or visit http://marketplace.aami.org.